. 2022 Aug 30;10:969829. doi: 10.3389/fpubh.2022.969829

Table 1.

Photon counts and standard deviation for the background and 200 mGy irradiated salt samples for five sets of measurements on three instruments, with calculated sensitivities and detection limits.

Instrument and wavelength
PPSL Portable OSL 1 Portable OSL 2
IR (890 nm) IR (940 nm) IR (890 nm) OSL (470 nm)
Background (empty chamber) 20 ± 34 −10 ± 32 −4 ± 31 45 ± 25 −61 ± 26
MDL (counts) 123 95 94 74 78
Pre-irradiation counts 113 ± 72 −4 ± 73 −146 ± 89 −27 ± 128 673 ± 160
200 mGy counts 279,200 ± 71,200 605,900 ± 60,900 65,500 ± 21,800 43,300 ± 1,700 230,3300 ± 80,700
Sensitivity (counts mGy−1) 1,400 ± 240 3,030 ± 300 330 ± 110 220 ± 10 11,520 ± 400
MDD (mGy) 0.088 0.031 0.286 0.340 0.0067