. 2022 Aug 30;10:969829. doi: 10.3389/fpubh.2022.969829

Table 2.

Mean and standard deviation of photon counts recorded using the SUERC portable OSL reader at 470 nm, the IRSL instrument at 890 nm and the pulsed PSL instrument, from sets of ten individual sachets of salt exposed to gamma doses from 0–500 μGy.

Portable OSL 470 nm Portable IRSL 890 nm Pulsed PSL 880 nm
Dose (μGy) Counts CV (%) Counts CV (%) Counts CV (%)
0 3,534 ± 453 12.8 1,204 ± 313 26.0 1,324 ± 102 7.7
20 3,307 ± 689 20.8 1,443 ± 300 20.8 1,256 ± 150 12.0
50 5,091 ± 460 9.0 1,843 ± 607 32.9 1,487 ± 122 8.2
100 7,323 ± 444 6.1 2,214 ± 202 9.1 1,701 ± 143 8.4
200 11,860 ± 1,086 9.2 2,388 ± 655 27.5 2,038 ± 237 11.7
500 25,202 ± 2,303 9.1 4,565 ± 687 15.1 3,649 ± 385 10.6