. 2022 Sep 2;12:964455. doi: 10.3389/fonc.2022.964455

Table 1.

Dosimetry evaluation of anthropomorphic phantom with 3D-printed bolus .

PTV Mean value OARs Mean value
*Dmin, cGy 4932 Ipsilateral lung Dmean, cGy 1017
Dmax, cGy 5259 V5 38.80%
Dmean, cGy 5131 V20 19.56%
D95%, cGy 5021 V30 12.54%
D2%, cGy 5251 Heart Dmean, cGy 438
D98%, cGy 4977 V30 1.42%
D50%, cGy 5134 V40 0.30%
HI 0.05 Spinal Cord Dmax, cGy 81
CI 99.91% Spinal Cord PRV Dmax, cGy 88

3D, three-dimensional; PTV, planning target volume; HI, homogeneity index ((D2%-D98%)/D50%); CI, conformity index; PRV, planning organs at risk volume.

*Dmin, minimum dose of the target volume; Dmax, maximum dose of the target volume; Dmean, mean dose of the target volume; D95%, the dose that covers 95% of the target volume.