Table 1.

Dosimetric comparison of scan protocols.

Scanner Voltage
[kVp]
Current–time product
[mAs]
CTDIvol
[mGy]

MDCT

Siemens

SOMATOM

Force

Effective per scan
70 50.0 1.3
70 97.0 2.7
80 122.0 5.1
70 199.0 5.3
Sn 100 898.0 7.5
Sn 150 146.0 8.2
100 113.0 9.8
80 241.0 10.2
120 101.0 14.1
140 96.0 18.9
100 227.0 19.5
80 489.0 20.3
120 202.0 28.1
140 192.0 37.7
100 453.0 39.3
120 405.0 54.0
140 383.0 75.2

CBCT

Siemens

MULTITOM

Rax

Mean per pulse
60 0.6 2.3
60 1.2 4.5
80 0.6 4.5
102 0.6 9.1
60 2.5 9.5
80 1.2 9.9
117 0.6 13.6
102 1.2 16.3
80 2.5 17.4
133 0.6 21.6
117 1.2 23.6
102 2.5 34.6
133 1.2 36.3
117 2.5 44.1
133 2.5 61.0

CTDIvol, computed tomography dose index calculated for a 16 cm phantom; CBCT, cone-beam computed tomography; MDCT, multidetector computed tomography; Sn, 0.4 mm tin prefiltration.