Inter-assay precision and accuracy for quality control (QC) of plasma samples spiked with antidepressant drugs. AMI, amitriptyline; MAP, maprotiline; VEN, venlafaxine; NOR, nortriptyline; IMI, imipramine; DES, desipramine for different methods applied. LLOQ lower limit of quantification, LQC low QC, MQC medium QC, HQC high QC

QC level Method AMI MAP VEN NOR IMI DES

LLOQ

(25 ng/mL)

LC-vDMS-MS (SIM) Accuracy (%) 102 91.8 104 100 107 97.3
Precision (%) 10.4 8.75 12.9 8.91 10.2 9.94
S/N 52 41 350 253 243 117
LC–MS (MRM) Accuracy (%) 110 100 114 101 110 108
Precision (%) 6.12 8.59 6.58 2.85 1.97 4.68
S/N 327 702 50 384 436 1637

LQC

(80 ng/mL)

LC-vDMS-MS (SIM) Accuracy (%) 100 95.4 93.4 99.3 106 86.1
Precision (%) 8.12 6.71 7.23 12.5 11.3 5.63
S/N 424 178 1492 694 622 367
LC–MS (MRM) Accuracy (%) 93.3 89.3 87.4 87.0 89.8 92.0
Precision (%) 5.94 4.84 4.28 9.45 7.13 3.67
S/N 594 1296 120 1096 2157 2343

MQC

(50 ng/mL)

LC-vDMS-MS (SIM) Accuracy (%) 98.5 93.9 94.8 93.3 93.6 98.4
Precision (%) 7.44 10.8 9.59 8.46 9.12 8.15
S/N 1964 448 6551 2135 2868 901
LC–MS (MRM) Accuracy (%) 99.5 108 92.3 96.6 111 104
Precision (%) 7.80 6.43 4.29 5.66 5.85 10.9
S/N 8018 12,012 402 8805 24,475 17,128

HQC

(1500 ng/mL)

LC-vDMS-MS (SIM) Accuracy (%) 101 94.1 93.7 96.5 95.3 92.0
Precision (%) 11.4 9.57 7.02 6.35 9.17 9.21
S/N 2797 898 6702 3585 3946 1499
LC–MS (MRM) Accuracy (%) 91.5 97.3 93.8 94.9 93.4 95.0
Precision (%) 9.29 6.15 9.29 7.33 4.94 6.39
S/N 17,696 18,303 860 17,219 41,603 34,722