Skip to main content
. 2022 Sep 6;13:925018. doi: 10.3389/fphar.2022.925018

TABLE 3.

Intra-batch and inter-batch precision and accuracy for SRL in cryopreserved and fresh human whole blood.

Matrix Intra-batch (n = 6) Inter-batch (n = 6 × 3)
LLOQ QC LQC MQC HQC LLOQ QC LQC MQC HQC
A P A P A P A P A P A P A P A P
Cryopreserved human whole blood −2.4 8.0 −0.7 12.8 −2.7 6.8 4.3 7.7 −4.0 8.8 −11.3 10.5 1.3 7.9 5.7 8.5
Fresh human whole blood 11.8 2.9 4.0 6.4 0.0 6.7 3.0 9.0 5.0 13.0 2.0 10.5 0.0 7.3 6.8 7.3

Note: A, accuracy and data are expressed as relative error (RE, %); P, precision, and data are expressed as the relative standard deviation (RSD, %); n, number of replicates; LLQC, 0.500 ng/ml; LQC, 1.50 ng/ml; MQC, 15.0 ng/ml; HQC, 40.0 ng/ml.