. 2022 Sep 20;6(6):e12772. doi: 10.1002/rth2.12772

TABLE 1.

D‐dimer and DDR results of 12 samples on four commercial assays

D‐dimer value D‐dimer ratio
Assay 1 a Assay 2 b Assay 3 c Assay 4 d Maximum difference e Assay 1 Assay 2 Assay 3 Assay 4 Maximum difference
ULN 243 0.5 1 0.5
Report unit ng/mL mg/L μg/ml μg/ml
Express unit DDU FEU DDU FEU
Sample 1 130 0.32 0.60 0.27 481.5 0.53 0.64 0.6 0.54 1.21
Sample 2 132 0.25 0.96 0.34 528.0 0.54 0.50 0.96 0.68 1.92
Sample 3 153 0.26 0.78 0.37 588.5 0.63 0.52 0.78 0.74 1.50
Sample 4 254 0.64 1.02 0.64 396.9 1.04 1.28 1.02 1.28 1.25
Sample 5 755 2.05 2.87 1.97 383.2 3.11 4.10 2.87 3.94 1.43
Sample 6 967 2.52 3.83 2.55 383.7 3.98 5.04 3.83 5.10 1.33
Sample 7 1030 1.82 3.70 1.59 647.8 4.24 3.64 3.70 3.18 1.33
Sample 8 1525 3.07 5.33 3.34 496.7 6.28 6.14 5.33 6.68 1.25
Sample 9 2091 4.67 5.97 4.01 521.4 8.61 9.34 5.97 8.02 1.56
Sample 10 2772 6.18 7.40 4.97 557.7 11.41 12.36 7.40 9.94 1.67
Sample 11 2891 5.83 8.07 5.94 495.9 11.90 11.66 8.07 11.88 1.47
Sample 12 5095 7.44 13.38 8.01 684.8 20.97 14.88 13.38 16.02 1.57
Total 513.8 ± 97.4 1.46 ± 0.21

Abbreviations: DDU, D‐dimer unit; FEU, fibrinogen equivalent unit; ULN, the upper limit of the normal range.

a

Assay 1, Werfen ACL‐TOP 750 with HemosIL D‐Dimer HS reagent;

b

Assay 2, Sysmex CS5100 with Siemens Innovance D‐dimer reagent;

c

Assay 3, Sekisui CP3000 with Nanopia D‐dimer reagent;

d

Assay 4, Stago STA‐R MAX with STA Liatest D‐Di reagent.

e

Maximum difference means that the maximum value of four results from the same sample divided by the minimum; e.g., Sample 12: 5095 divided by 7.44 is 684.8 for D‐dimer value, and 20.97 divided by 13.38 is 1.57 for D‐dimer ratio.