TABLE 5.

Performance of molecular methods (compared with the gold standard reference method) currently in clinical use for H. pylori detection, according to recent reportsa

Test Gold standard method Sensitivity (%) Specificity (%) PPV (%) NPV (%) Accuracy (%) Reference
Nested PCR UBT 93.8 95.9 299
Multiplex PCR CIM and CLO 100 59.6 73.8 100 81.1 300
BACTfish H. pylori Combi test (Izinta Kft., Budapest, Hungary) Histology 98 99.6 98.9 99.2 250
RIDA GENE H. pylori (R-Biopharm AG, Darmstadt, Germany) Culture alone or histopathology and RUT combined 100 99 94 100 318
Amplidiag H. pylori+ClariR (Mobidiag, Espoo, Finland) Culture or TaqMan RT-PCR 96.3 98.7 92.2 99.3 319
MutaREAL H pylori (Inmundiagnostik, Bensheim, Germany) Culture 93.3 86.9 90.3 90.9 320
ViaSure H. pylori real-time PCR detection kit (CerTest Biotec S.L., Zaragoza, Spain) SAT 85.7 100 100 84.6 92 321
a

PPV, positive predictive value; NPV, negative predictive value.