Table 2.
Method Qualification Data for Total Analytes in Male SD Rat Plasma
Compound | Concentration Range (ng/mL) | LODa (ng/mL) | LOQb (ng/mL) | Standard c | QCd | ||
---|---|---|---|---|---|---|---|
Intra-day Accuracy (RE, %)e | Intra-day Precision (RSD, %)f | Intra-day Accuracy (RE, %)e | Intra-day Precision (RSD, %)f | ||||
P-BZT | 10–1000 | 1.95 | 10.1 | −14.3 to 5.9 | 1.4 to 18.0 | −5.1 to −1.5 | 8.8 to 15.1 |
Octrizole | 5–100 | 1.53 | 5.0 | −7.5 to 11.4 | 0.8 to 9.1 | −45.5 to 2.4 | 8.6 to 27.8 |
Drometrizole | 1–100 | 0.14 | 1.0 | −8.8 to 4.6 | 0.4 to 5.0 | −25.2 to −3.6 | 15.8 to 17.3 |
tBuPrOcEst-BZT | 1–100 | 0.12 | 1.0g | −7.1 to 4.5 g | 0.0 to 7.0 | −12.2 to 0.7g, −27.3 to −20.5h | 6.7 to 10.2g 2.9 to 6.0h |
Bumetrizole | 1–100 | NA | 1.0 | −8.1 to 17.8 | 5.5 to 7.3 | −17.0 to 0.5 | 23.1 to 28.1 |
DiMeEtPh-BZT | 1–100 | 0.22 | 1.0 | −11.5 to 14.0 | 2.7 to 17.1 | 3.0 to 4.4 | 9.3 to 25.7 |
ditBuCl-BZT | 1–100 | 0.04 | 1.0 | −3.4 to 3.0 | 1.0 to 8.2 | 8.5 to 3.2 | 5.6 to 10.0 |
DitPe-BZT | 5–500 | NA | 5.0 | −18.2 to 17.2 | 5.0 to 12.9 | 8.0 to 12.7 | 4.2 to 8.4 |
tBu-BZT | 1–100 | NA | 1.0 | −13.2 to 11.9 | 1.6 to 12.4 | 2.4 to 4.0 | 8.4 to 23.9 |
LOD= Limit of Detection; 3×Standard Deviation for 6 lower limit of quantitation (LOQ) replicates
LOQ=lower limit if quantitation; is the lowest concentration of the calibration curve
Estimated based on plasma calibration standards, n = 2
Estimated based on low and high-level Quality Control (QC) samples, n = 4; 20 and 800 ng/mL for P-BZT; 9 and 75 ng/mL for Octrizole, 2 and 50 ng/mL for drometrizole, bumetrizole, DiMeEtPh-BZT, and ditBu-BZT; 10 and 250 ng/mL for DitPe-BZT; and 1.5 and 80 ng/mL for total tBuPrOcEst-BZT (as tBuPrOH-BZT))
RE %= percent relative error
%RSD = percent relative deviation
With respect to tBuPrOH-BZT
With respect to tBuPrOcEst-BZT