| Binsch et al [21], 2016 |
Resilience and workload monitoring |
PPGa reliable in the at-rest condition; wide-ranging outcomes during movement
Apple Watch showed the most variance in steps and distances compared with ground truth measurements, followed by the comparison, Fitbit Surge and Microsoft Band
Such variances are surmised to be because of differences in data resolution and access and underlying algorithms using accelerometer and GPS data for step count estimation
|
| Shcherbina et al [22], 2017 |
HRb and EEc
|
Lowest error in HR and EE for cycling; highest error for walking
Apple Watch achieved the lowest overall error in HR and EE of the tested devices (Basis Peak, Fitbit Surge, Microsoft Band, Mio Alpha 2, PulseOn, and Samsung Gear S2)
|
| Dooley et al [23], 2017 |
HR and EE |
Apple Watch HR mean absolute percentage error was between 1.14% and 6.70%, not significantly different during baseline and vigorous-intensity treadmill exercise; lower HR in light- or moderate-intensity treadmill exercise and recovery
EE mean absolute percentage error was between 14.07% and 210.84%, measuring higher EE in all states compared with the criterion measure (Parvo Medics TrueOne 2400), with greater errors for higher BMI and the male population
HR and EE results were mostly better than other tested devices (Fitbit Charge HR and Garmin Forerunner 225)
|
| Wang et al [24], 2017 |
HR |
|
| Hernando et al [25], 2018 |
HRVe
|
Apple Watch RR interval data were found to contain gaps lasting 6.5 seconds per gap, averaging 5 gaps per recording, not correlated with stress or relaxation case
The cause is surmised to be because of failure to detect reliable pulses from PPG data
Temporal HRV indices were not significantly affected, but frequency-based LFf and HFg power showed a significant decrease
Apple Watch was able to successfully detect mild mental stress
|
| Abt et al [26], 2018 |
Moderate-intensity exercise |
|
| Abt et al [27], 2018 |
Maximal HR |
Apple Watch had good to very good criterion validity for measuring maximal HR with no substantial under- or overestimation
Moderate and small errors were found for simultaneous recording from left versus right watches
|
| Roomkham et al [28], 2019 |
Sleep monitoring |
Apple Watch raw acceleration data were used to compute ENMOi for classification
Apple Watch had high accuracy (97.3%) and sensitivity (99.1%) in detecting sleep and adequate specificity (75.8%) in detecting wakefulness
|
| Perez et al [29], 2019 |
AFj
|
Apple Watch irregular rhythm notification was triggered on 0.52% of 419,297 participants
Of those who returned an ECGk patch, 84% of subsequent notifications were confirmed to be AF
A total of 34% of ECG patches returned identified AF in part because of the transient nature, suggesting that Apple Watch may be useful for ongoing monitoring
|
| Nuss et al [30], 2019 |
EE |
Apple Watch overestimated EE in women and underestimated EE in men
Pooled relative error was 24.3%, 18.6% for men, and 19.9% for women
Neither device showed accurate results compared with EE measured with a MetCart
|
| Thomson et al [31], 2019 |
HR |
ECG correlation was strongest for very light intensity with a >0.90 concordance correlation coefficient
Most relative error rates were <5% with a maximum of 5.73%
Apple Watch was more accurate in recording HR than the Fitbit Charge HR 2
|
| Nelson and Allen [32], 2019 |
HR and passive monitoring |
Apple Watch 3 was generally accurate across a 24-hour period compared with ECG; the mean difference was −1.8 bpm, the mean absolute error was 5.86%, and the mean agreement was 95%
Apple Watch was more accurate than Fitbit Charge 2
|
| Falter et al [33], 2019 |
HR and EE in patients with cardiovascular disease |
Apple Watch showed good correlation without systematic error comparing Apple Watch PPG HR with ECG ground truth
Apple Watch showed a systematic overestimation of EE compared with indirect calorimetry
Apple Watch HR accuracy was clinically acceptable
|
| Düking et al [34], 2020 |
HR and EE |
Apple Watch 4 showed the highest validity in measuring HR, followed by Polar Vantage V, Garmin Fenix 5, and Fitbit Versa
The coefficient of variation for HR was 0.9% to 4.3% and, for EE, it was 13.5% to 27.1%
|
| Espinosa et al [35], 2020 |
Step counting and HR |
The walking error was 2.6%; jogging error was 5.1%
HR limit of agreement was −2.2 to 1.8 bpm for walking and −3.5 to 4.3 bpm for jogging
Apple Watch displayed a high level of agreement and was highly accurate
|
| Seshadri et al [36], 2020 |
HR in patients with AF |
Patients with AF showed a correlation coefficient of 0.7 between Apple Watch 4 and telemetry
Apple Watch 4 HR was more accurate for patients in the AF condition than for those not in the AF condition
Caution suggested in Apple Watch HR monitoring in patients with arrhythmia
|
| Seshadri et al [37], 2020 |
AF |
Apple Watch 4 notification correctly identified AF in 34 of 90 instances (41% sensitivity), with no false positives and 31% inconclusive
The agreement between Apple Watch 4 and telemetry was 61%
Apple Watch–exported ECG PDF files showed AF in 84 of 90 instances (96% sensitivity), no false positives, and 2 failures to generate PDFs
Agreement between Apple Watch 4 ECG PDFs and telemetry was 98.9%
Further validation is required because of the high inconclusive result rate
|
| Glasheen et al [38], 2021 |
Wheelchair use |
Apple Watch 1 only showed good agreement on higher-rate fixed-frequency tasks, with significant overestimation at low frequency
Arm ergometry showed good agreement across all cadences
Overground tasks showed poor agreement, with significant differences found
|
| Huynh et al [39], 2021 |
HR in patients with obstructive sleep apnea and AF |
Apple Watch 1 variability increased as the magnitude of the HR measurement increased
The Lin concordance correlation coefficient was 0.88, suggesting acceptable agreement between Apple Watch 1 and telemetry
|