Table 1. Ion Accumulation Times (IT) and TIC Values for 20 Consecutive Scans of a Sample of Rat Brain Extract of 0.086 mg/mL Acquired with DIP, ESI, and HESI Sourcesa.

  IT (ms) ± RSD (%) TIC ± RSD (%)
DIP 181.15 ± 3.02 1.17 × 107 ± 3.78
ESI 1.41 ± 8.61 1.54 × 109 ± 7.71
HESI 1.17 ± 6.30 1.84 × 109 ± 5.93
a

The flow rate of DIP is 1 μL/min and the flow rate of ESI and HESI is 5 μL/min.