Table 1. Ion Accumulation Times (IT) and TIC Values for 20 Consecutive Scans of a Sample of Rat Brain Extract of 0.086 mg/mL Acquired with DIP, ESI, and HESI Sourcesa.
IT (ms) ± RSD (%) | TIC ± RSD (%) | |
---|---|---|
DIP | 181.15 ± 3.02 | 1.17 × 107 ± 3.78 |
ESI | 1.41 ± 8.61 | 1.54 × 109 ± 7.71 |
HESI | 1.17 ± 6.30 | 1.84 × 109 ± 5.93 |
The flow rate of DIP is 1 μL/min and the flow rate of ESI and HESI is 5 μL/min.