Skip to main content
. 2022 Sep 22;12:15816. doi: 10.1038/s41598-022-20222-7

Table 3.

CFRP screw accuracy and assessability depending on imaging modality.

iCT rCBCT CBCT p-value
Total no. of navigated screws 93 99 51 n/a
Not assessable, n (%) 1 (1%, T8) 9 (9%, T 7–12) 3 (6%, T11–12)

*0.0472

(iCT vs. rCBCT)

Screws intraoperatively corrected, n (%) 6 (7%) 0 0 0.0853
Screw accuracy (A + B), n (%) 69 (74%) 68 (69%) 25 (49%)

**0.0064

(rCBCT vs. CBCT and iCT vs. CBCT)

Pedicle breach 2–4 mm (C), n (%) 14 (15%) 19 (19%) 13 (25%) 0.3395
Pedicle breach > 4 mm (D), n (%) 9 (10%) 3 (3%) 10 (20%)

**0.0042

(rCBCT vs. CBCT)

Median pedicle isthmus diameter of instrumented pedicles without breach (A, mm) 6.5 (5.5–13) 6.5 (5.5–15) 9 (6–12)

*0.0407

(rCBCT vs. CBCT)

Median pedicle isthmus diameter of instrumented pedicles with breached screws (B–D, mm) 5.7 (3.9–8.8) 5.8 (3.4–10) 6.1 (4.5–11) 0.4982
Median isthmus/screw diameter–ratio in screws without breach (A) 1.2 1.2 1.4 0.1180
Median isthmus/screw diameter–ratio in breached screws (B–D) 1.04 1.1 1.1 0.9360
Number of pedicle screws larger than pedicle isthmus, n (%) 5 (5%) 13 (13%) 10 (19%)

*0.0353

(iCT vs. CBCT)

Screw accuracy (A + B) without screws larger than the pedicle isthmus, n (%)

67/91 (74%)

Th: 39/57 (68%) TLJ: 24/26 (92%) L: 4/4 (100%)

(n.s.)

56/77 (73%)

Th: 41/60 (68%) TLJ: 8/10 (80%) L: 6/6 (100%)

(n.s.)

17/37 (46%)

Th: 5/20 (25%) TLJ: 5/8 (63%) L: 7/8 (88%)

(*p = 0.0202)

**0.0056

(iCT vs. CBCT, rCBCT vs. CBCT)

Median time for accuracy assessment (seconds) 120 (27–540) 120 (17–720) 180 (24–1080) 0.5551

Values are given in total number with percentage or as median with total range, as appropriate.

iCT Intraoperative CT, rCBCT robotic cone beam CT, CBCT cone beam CT, n.s. not statistically significant. Based on SIN score: L = lumbar (L2-5), Th = thoracic (Th1-10), TLJ = thoraco-lumbar junction (Th11-L1).