TABLE 7.
Heterogeneous TPS photon beam validation checks
| # | Objective | Examples | Tolerances a | References |
|---|---|---|---|---|
| 6.1 | Validate planning system reported electron (or mass) densities against known values | CT‐density calibration for air, lung, water, dense bone, and possibly additional tissue types | – | TG‐65, 47 IAEA TRS‐430 3 |
| 6.2 | Heterogeneity correction distal to lung tissue | 5 × 5 cm2, measure and calculate dose ratio above and below heterogeneity, outside of the buildup region | 3% | IAEA TRS‐430, 3 Carrasco et al. 43 |
Tolerances are relative to local dose unless otherwise noted.