TABLE 10.
Basic TPS validation tests for electron beams and minimum tolerance values
| # | Dose comparison objective | Examples | Tolerances |
|---|---|---|---|
| 8.1 | Verify that TPS reproduces the absolute dose under reference calibration conditions | Calculate the absolute dose per MU to the reference point using the same SSD and cone size as the calibration protocol | 0.5% |
| 8.2 | Basic model verification with shaped fields | PDD and profile comparison of custom cutouts at standard and extended SSDs | 3%/3 mm |
| 8.3 | Surface irregularities/obliquity | PDD and/or profile comparison of obliquely incident fields using reference cone and nominal clinical SSD | 5% |
| 8.4 | Inhomogeneity test | Compare manual dose calculation or chamber measurement to TPS with reference cone and nominal clinical SSD | 7% |