TABLE 10.

Basic TPS validation tests for electron beams and minimum tolerance values

# Dose comparison objective Examples Tolerances
8.1 Verify that TPS reproduces the absolute dose under reference calibration conditions Calculate the absolute dose per MU to the reference point using the same SSD and cone size as the calibration protocol 0.5%
8.2 Basic model verification with shaped fields PDD and profile comparison of custom cutouts at standard and extended SSDs 3%/3 mm
8.3 Surface irregularities/obliquity PDD and/or profile comparison of obliquely incident fields using reference cone and nominal clinical SSD 5%
8.4 Inhomogeneity test Compare manual dose calculation or chamber measurement to TPS with reference cone and nominal clinical SSD 7%