Table 1.
MagnifEye label |
||||
---|---|---|---|---|
Positive | Negative | Void | ||
ATS label (Original not clinically assessed) | Positive | 93 | 8 | 0 |
Negative | 85 | 56,463 | 67 | |
Void | 1 | 25 | 34 |
95% confidence interval (CI) | ||
---|---|---|
Sensitivity | 92.08% | 84.99–96.52% |
Specificity | 99.85% | 99.81%–99.88% |
Positive predictive value | 52.25% | 46.75–57.69% |
Negative predictive value | 99.99% | 99.97–99.99% |
Void accuracy | ||
Accuracy | 99.84% | 99.80–99.87% |