. 2022 Sep 27;3(10):100784. doi: 10.1016/j.xcrm.2022.100784

Table 1.

Confusion matrix for cases that passed QC, showing ATS and MagnifEye labels (substudy 1)

MagnifEye label
Positive Negative Void
ATS label (Original not clinically assessed) Positive 93 8 0
Negative 85 56,463 67
Void 1 25 34
95% confidence interval (CI)
Sensitivity 92.08% 84.99–96.52%
Specificity 99.85% 99.81%–99.88%
Positive predictive value 52.25% 46.75–57.69%
Negative predictive value 99.99% 99.97–99.99%
Void accuracy
Accuracy 99.84% 99.80–99.87%