Table 5.
Confusion matrix showing the “ground truth” result and the user read (substudy 2, excluding invalid images)
User read |
||||
---|---|---|---|---|
Positive | Negative | Void | ||
Ground truth (clinically assessed) | positive | 2 | 12 | 0 |
negative | 29 | 33,193 | 2 | |
void | 0 | 3 | 13 |
95% CI | ||
---|---|---|
Sensitivity | 6.45% | 0.79%–21.42% |
Specificity | 99.96% | 99.94%–99.98% |
Positive predictive value | 14.29% | 0.18%–42.81% |
Negative predictive value | 99.91% | 99.87%–99.94% |
Void accuracy | 81.25% | 54.35%–95.95% |
Accuracy | 99.86% | 99.82%–99.90% |