Validation of the Detection repeatability of the developed triplex real-time PCR method.

Genes DNA (copies/μL) Rounds of testing Ct meana Ct Standard deviation C.V.a C.V. between groups
B646L 7.89 × 102 1 33.34 0.24 0.73% 0.74%
2 32.99 0.26 0.79%
3 33.47 0.42 1.25%
7.89 × 103 1 29.53 0.11 0.37% 0.49%
2 29.33 0.16 0.54%
3 29.60 0.21 0.70%
7.89 × 104 1 26.68 0.09 0.33% 0.17%
2 26.59 0.08 0.29%
3 26.63 0.05 0.19%
7.89 × 105 1 23.14 0.12 0.54% 0.19%
2 23.16 0.09 0.40%
3 23.07 0.02 0.11%
7.89 × 106 1 19.15 0.13 0.67% 0.54%
2 18.95 0.15 0.77%
3 19.02 0.09 0.46%
MGF_360-14L 4.70 × 102 1 35.29 0.66 1.86% 1.35%
2 34.41 1.40 4.06%
3 34.58 0.20 0.56%
4.70 × 103 1 30.53 0.06 0.18% 0.36%
2 30.62 0.05 0.16%
3 30.41 0.09 0.30%
4.70 × 104 1 26.06 0.19 0.71% 0.28%
2 25.96 0.23 0.90%
3 26.10 0.06 0.24%
4.70 × 105 1 22.46 0.04 0.16% 0.56%
2 22.49 0.09 0.41%
3 22.69 0.18 0.78%
4.70 × 106 1 18.49 0.05 0.28% 0.14%
2 18.44 0.13 0.72%
3 18.48 0.05 0.28%
CD2v 8.21 × 102 1 32.18 0.16 0.51% 0.29%
2 32.03 0.20 0.63%
3 32.01 0.36 1.11%
8.21 × 103 1 29.40 0.07 0.25% 0.31%
2 29.39 0.05 0.17%
3 29.55 0.11 0.36%
8.21 × 104 1 26.48 0.17 0.64% 0.23%
2 26.43 0.30 1.15%
3 26.36 0.42 1.61%
8.21 × 105 1 23.78 0.06 0.27% 0.01%
2 23.78 0.04 0.19%
3 23.78 0.06 0.25%
8.21 × 106 1 19.76 0.07 0.35% 0.18%
2 19.69 0.02 0.10%
3 19.75 0.10 0.53%
a