. 2022 May 13;59(10):4097–4107. doi: 10.1007/s13197-022-05462-9

Table 1.

Optimized GC-MS parameter settings for analysis of CS2

Parameter Settings
GC parameters
Injector temperature 150 °C
Split injection Split ratio 20:1 (split flow 22.066 ml/min)
Carrier flow He, 1.1 ml/min
Injector conditions Temperature program: 70 °C hold 0.1 min, ramp at 450 °C/min to 325 °C, hold 2 min, cool at 10 °C/min to 250 °C.
Injection volume / mode 2 ml / split ratio 20:1
Column conditions (run) Temperature program of 14 min: 40 °C hold 5 min, ramp at 40 °C / min to 200 and hold for 5 min
Column conditions (post run) Post-run program of 10 min at 310 °C: Mid-column back flush, with inlet pressure at 2 psi, backflush column flow − 2.553 ml/min, and onward column flow 2.967 ml/min.
MS parameters
Source temperature 230 °C
Ionization / electron energy Electron ionization (EI) / 70 eV
Detector voltage 1500 V
Ion source temperature 230 °C
Damping gas flow 0.6 ml/min
Emission current 250 µA
MS analysis Selected ion monitoring (SIM), m/z 76 and 78 with unit resolution and dwell time 200 ms