Table 1. Rietveld analysis parameters obtained for LaB6 and the mixture of polycrystalline NaCl and sodalite.
The Rietveld refinement was done using the FPA. The fitted parameters were the background (simulated with a Chebyshev polynomial with seven fittable parameters) and a preferred orientation according to March. Fixed parameters determined on an LaB6 reference compound were ‘tube tails’ (a simple axial model for describing the axial divergence of the beam) and Lorentz polarization.
| Calibrant | Quantified mixture | ||
|---|---|---|---|
| Parameter | LaB6 | NaCl | Sodalite |
| Data collection | |||
| Wavelength (Å) | 1.540562 | 1.540562 | 1.540562 |
| Monochromator | Ge(111) | Ge(111) | Ge(111) |
| No of scans averaged† | Up to 24 | 7 | 7 |
| Exposure time per scan (mins) | 12 | 15 | 15 |
| Digitized step size (°) | 0.012 | 0.012 | 0.012 |
| 2θ range used in RR‡ (°) | 10–90 | 10–90 | 10–90 |
| Unit-cell and refined parameters | |||
| Space group |
|
|
|
| a (Å) | 4.15685 (9) | 5.63982 (2) | 8.87640 (4) |
| Cell volume (Å3) | 71.828 | 179.389 (3) | 699.375 (12) |
| Density (g cm−3) | 4.7108 (4) | 2.1640 | 2.3012 |
| Lattice strain ɛ (°) | 0.000032 (2) | 0.00435 (1) | 0.00109 (1) |
| Residual factors | |||
R
p,
(%) |
5.2, 16.9 | 4.2, 24.6 | |
| Sample composition | |||
| Purity of material (%) | 99.9 | >99 | >99 |
| Relative quantity weighed (%) | 58.5 ± 0.3 | 41.5 ± 0.3 | |
| Relative quantity obtained from RR‡ (%) | 60.0 ± 0.1 | 39.9 ± 0.1 | |
For LaB6, the sample was rotated by a few degrees after each exposure, while the mixture of NaCl and sodalite on the same sample holder was repeatedly irradiated keeping the same orientation of the sample holder.
RR = Rietveld refinement.