. 2022 May 12;24(7):1370–1379. doi: 10.1111/dom.14714

TABLE 1.

MiniMed 780G system use and settings post‐AHCL initiation (cohort 1)

All users Users ≤15 y Users >15 y P a
Users, n 12 870 3211 8874
Sensor use, % 90.5 ± 10.2 90.5 ± 10.0 90.5 ± 10.2 .271
Time in AHCL, % 92.3 ± 13.0 92.7 ± 12.3 92.3 ± 13.0 .260
SMBG measurements per day, n 3.4 ± 1.1 3.6 ± 1.1 3.4 ± 1.0 <.0001
AHCL exits per week, n
Total 1.0 ± 1.1 1.1 ± 1.2 1.0 ± 1.0 <.0001
Initiated by the user 0.5 ± 1.0 0.6 ± 1.1 0.5 ± 0.9 .595
Initiated by the system 0.5 ± 0.6 0.6 ± 0.7 0.5 ± 0.6 <.0001
Glucose target, % of time
100 mg/dl (5.6 mmol/L) 47.7 ± 43.9 45.1 ± 43.6 48.7 ± 43.9 .0008
110 mg/dl (6.1 mmol/L) 23.2 ± 36.0 25.7 ± 37.1 22.5 ± 35.7 <.0001
120 mg/dl (6.7 mmol/L) 22.3 ± 36.7 22.2 ± 36.1 22.1 ± 36.7 .883
150 mg/dl (8.3 mmol/L) 1.9 ± 4.5 2.2 ± 4.9 1.9 ± 4.4 .002
Open loop 4.8 ± 9.2 4.8 ± 8.7 4.8 ± 9.2 .961
Active insulin time, % of time
2 h 37.6 ± 45.7 45.5 ± 47.2 35.5 ± 45 <.0001
>2 to 3 h 53.1 ± 46.5 48.7 ± 46.7 54.2 ± 46.4 <.0001
>3 to 4 h 8.9 ± 26.5 5.4 ± 20.7 10.0 ± 27.9 <.0001
>4 h 0.4 ± 5.3 0.3 ± 4.7 0.3 ± 5.1 .617

Note. All values are shown as mean ± standard deviation.

Abbreviations: AHCL, advanced hybrid closed loop; SMBG, self‐monitored blood glucose.

a

Comparison between users aged ≤15 and >15 years.