Table 2.
The ability of thyroid function test to detect deviations from the euthyroid state among euthyroid participants and patients with subclinical (mild) hypothyroidism. Calculations are based on the measurements listed in Table 1
| TSH | TT4 | fT4 index | fT4 | |
|---|---|---|---|---|
| Laboratory reference range: | 0.3–4.5 mIU/L | 60–140 nM | 70–140 nM | 12–22 pM |
| Samples within the reference range: | ||||
| Euthyroid, % (n) | 100 (180) | 99.4 (179) | 97.8 (176) | |
| Mildly hypothyroid, % (n)a | 12.9 (31) | 97.1 (233) | 78.3 (188) | |
| Sensitivity, % | 87.1 | 2.9 | 21.7 | |
| Specificity, % | 100 | 99.4 | 98.9 | |
| PPV, positive predictive value, % | 100 | 87.5 | 96.3 | |
| NPV, negative predictive value, % | 85.3 | 43.4 | 48.8 | |
TSH Thyrotropin, TT4 Total thyroxine, fT4 Free thyroxine
a One participant had subsequent measures of TSH and T4 within the reference range. Without this participant, 8.3% of TSH and 96.9% of T4 measurements were within the population-based reference range in this group