1 |
Verification |
|
What is (expected) numerical error for a new patient in the intended patient population given the chosen mesh resolution? |
2 |
Verification |
|
If there is a remote operator who will do some manual tasks (e.g., in image segmentation or running simulations), what are the potential errors due to intra- or inter-operator variability? |
3 |
Validation |
|
If validation involved a clinical study with an intermediate QOI (see text)–what is the relationship between the intermediate QOI and COU QOI? |
4 |
Validation |
|
If ‘personalized model validation’, that is, validation for each new patient—how relevant are the validation QOIs to the COU QOIs? |
5 |
UQ–model form |
Tissue model |
Is model form appropriate for all new patients in the intended patient population? Are there sub-populations the tissue model may not adequately represent? |
6 |
UQ–model form |
Cell model |
Is the cell model appropriate for all patients in intended patient population? Are there sub-populations the cell model may not adequately represent? |
7 |
UQ–personalized parameters |
Heart shape |
What is the potential error in the personalized heart shape for a new patient and what is the resultant impact on the tool outputs? |
8 |
UQ–personalized parameters |
BZ & scar |
What is the potential error in the personalized BZ/scar for a new patient and what is the resultant impact on the tool outputs? |
10 |
UQ–personalized parameters |
Fibers |
If a rule-based method was used—what is the potential error in specified fibers for a new patient and what is the impact on tool outputs? |
11 |
UQ–personalized parameters |
Cell model |
For personalized cell model parameters—what is the potential error in the personalized values of these parameters for a new patient and what is the resultant impact on the tool outputs? |
12 |
UQ–non- personalized parameters |
Conductivity |
If a fixed (not personalized) conductivity is used—are the tool outputs insensitive to population variability or other uncertainties in conductivity? |
13 |
UQ–non-personalized parameters |
Cell model |
For fixed (not personalized) cell model parameters—are the tool outputs insensitive to population variability or other uncertainties in these parameters? |