Concordance rate of bacteria’s Gram pattern in different bacteria concentrations by UF-5000.

Group Gram (+)/Gram(-) Mixed Undetectable overall
All patients (n=383)
Negative bacteriuria a 5.9% 0.0% 95.1% 76.9%
Positive bacteriuria b 38.3% 7.9% 100.0% 43.7%
Overall 33.9% 4.6% 95.1% 59.0%
Without mixed 70.5%
Patients without having antibiotic treatments 2 weeks prior to inclusion (n=269)
Negative bacteriuria a 8.3% 0.0% 96.2% 88.8%
Positive bacteriuria b 44.7% 3.7% 100.0% 30.3%
Overall 39.8% 2.3% 96.2% 65.4%
Without mixed 78.1%
Patients having antibiotic treatments 2 weeks prior to inclusion (n=104)
Negative bacteriuria a 0.0% 0.0% 92.0% 71.9%
Positive bacteriuria b 18.5% 20.0% 100.0% 18.9%
Overall 21.2% 14.3% 92.0% 52.5%
Without mixed 63.9%

MSU, midstream urine culture.

a

Negative bacteriuria: bacteria count by UF-5000<196/μl.

b

Positive bacteriuria: bacteria count by UF-5000≥196/μl.

The cutoff points of the BACT parameter (196/μl) were determined by the Youden index.