Skip to main content
. 2022 Oct 11;8:20552076221132127. doi: 10.1177/20552076221132127

Figure 1.

Figure 1.

Differences between simultaneous SpO2 readings of the smartwatch (Apple Watch 6) and oximeter (Masimo Radical-7) across different ranges of oxyhemoglobin saturation. Pooled SpO2 measurements were analyzed for all participants grouped. The solid line is the mean difference of the measurements (bias). Dashed lines are the 95% limits of agreement. The area of markers is proportional to the number of measurements.