. 2022 Oct 11;8:20552076221132127. doi: 10.1177/20552076221132127

Table 1.

Comparison of measurement bias and agreement.

SpO2a, % Biasb (95% CI), % Lower LOA (95% CI), % Upper LOA (95% CI), % Arms, %
Entire range 0.0 (−0.2 to 0.3) −5.8 (−6.2 to −5.4) 5.9 (5.5–6.3) 3.0
<90 1.2 (0.7 to 1.7) −5.3 (−6.1 to −4.4) 7.6 (6.7–8.4) 3.4
90–100 −0.3 (−0.6 to 0.1) −5.8 (−6.2 to −5.4) 5.1 (4.7–5.5) 2.8
a

[(smartwatch + oximeter)/2].

b

[smartwatch − oximeter].

LOA: 95% limits of agreement.