Skip to main content
. 2022 Sep 29;10:971096. doi: 10.3389/fbioe.2022.971096

TABLE 7.

Tibia plate mean RMSE and mean maximum OUH results for ground truth (GT) best sizes and predictions, split by manufacturer model sizings.

Model (no. Sizes) Mean GT best RMSE (mm) Mean prediction RMSE (mm) Mean GT best max OUH (mm) Mean prediction max OUH (mm)
Zimmer Biomet (6) 1.73 1.77 3.48 3.63
DePuy (8) 1.11 1.19 2.52 2.70
Smith & Nephew (8) 1.11 1.25 2.53 2.85
Maxx Orthopedics (8) 1.31 1.38 2.91 3.20
Stryker (8) 1.10 1.21 2.48 2.71
Mean 1.27 1.36 2.78 3.02