Table 2.
Full energy peak efficiencies of 2 × 2 inch LaBr3(Ce) scintillation detector.
Energy (keV) | This Study |
Kuluozturk and Demir |
Casanovas et al. |
Casanovas et al. |
Mouthi et al. |
|||||
---|---|---|---|---|---|---|---|---|---|---|
MCNPX MC simulation |
FLUKA MC simulation |
EGSnrc MC simulation |
Experimental |
Experimental |
||||||
Source-Detector distance |
Source-Detector distance |
Source-Detector distance |
Source-Detector distance |
Source-Detector distance |
||||||
2 cm | 5 cm | 2 cm | 5 cm | 2 cm | 5 cm | 2 cm | 5 cm | 2 cm | 5 cm | |
30.973 | 7.07 | 3.14 | 7 | 3.07 | - | - | - | - | - | - |
59.54 | 10.36 | 3.81 | 10.28 | 3.76 | - | 3.09 | - | 3 | - | 3.95 |
80.998 | 8.01 | 3.41 | 7.94 | 3.38 | - | 3.576 | - | 3.8 | - | - |
302.85 | 5.24 | 2.30 | 5.19 | 2.24 | - | 2.01 | - | 2 | - | 2.34 |
356.01 | 4.58 | 1.98 | 4.53 | 1.96 | - | 2.01 | - | 1.8 | - | 2.05 |
661.65 | 4.14 | 1.44 | 4.08 | 1.39 | - | 1.31 | - | 1.3 | - | 1.45 |
1173.24 | 2.35 | 0.87 | 2.38 | 0.83 | 2.15 | - | 2.1 | - | 2.48 | - |
1332.5 | 2.18 | 0.76 | 2.13 | 0.74 | 1.90 | - | 2 | - | 2.2 | - |
1408.01 | 2.07 | 0.72 | 2.02 | 0.70 | - | - | - | - | - | - |