Skip to main content
. 2021 Mar 6;1(1):100005. doi: 10.1016/j.xops.2021.100005

Table 2.

Interobserver, Interdevice, and Interocular Agreements in Definitely Decreased Autofluorescence Area and Square Root Area

Comparisons Sample Size Mean Difference Limits of Agreement P Value
DDAF area (mm2)
 Interobserver
 F.K.C. vs. R.C.H.J. baseline RE 61 –0.12 –3.13 to +2.87 0.528
 F.K.C. vs. R.C.H.J. baseline LE 59 –0.46 –4.68 to +3.73 0.104
 Interdevice
 O200 vs. H30 RE only 45 –0.19 –2.38 to +1.99 0.266
 H55 vs. O200 RE only 51 +2.79 –6.84 to +12.37 < 0.001
 H55 vs. H30 RE only 44 +1.20 –2.26 to +4.65 < 0.001
 Interocular
 RE vs. LE baseline 59 –2.24 –30.41 to +25.79 0.236
 RE vs. LE final 50 –2.03 –22.75 to +18.58 0.180
DDAF SRA (mm)
 Interobserver
 F.K.C. vs. R.C.H.J. baseline RE 61 –0.02 –0.20 to +0.17 0.168
 F.K.C. vs. R.C.H.J. baseline LE 59 –0.02 –0.31 to +0.26 0.197
 Interdevice
 O200 vs. H30 RE only 45 –0.03 –0.37 to +0.30 0.221
 H55 vs. O200 RE only 51 +0.27 –0.41 to +0.91 < 0.001
 H55 vs. H30 RE only 44 +0.16 –0.17 to +0.49 < 0.001
 Interocular
 RE vs. LE baseline 59 –0.11 –1.82 to +1.60 0.355
 RE vs. LE final 50 –0.13 –2.03 to +1.75 0.330

DDAF = definitely decreased autofluorescence; LE = left eye; H30 = Heidelberg 30° × 30° lens; H55 = Heidelberg 55° × 55° lens; O200 = Optos200 device; RE = right eye; SRA = square root area.

Bland-Altman 95% limits of agreement defined by mean ± 1.95 × standard deviation.

Paired sample t test.

Author initials.