Table 3.
Subgroup Analysis of Experimental Conditions versus Real-World Conditions in Agreement and Correlation between the NYU Langone Eye Test Application and the Rosenbaum Card
| Device by Conditions | Agreement with Rosenbaum Card (Mean ± 95% Limits of Agreement, logMAR) | Correlation with Rosenbaum Card (Intraclass Correlation Coefficient) |
|---|---|---|
| Experimental conditions (50 participants) | ||
| iPhone | 0.008 ± 0.16 | 0.90 |
| Android | 0.001 ± 0.13 | 0.92 |
| Real-world conditions (75 participants) | ||
| iPhone | 0.024 ± 0.35 | 0.66 |
| Android | 0.016 ± 0.36 | 0.63 |