Skip to main content
. 2022 Jun 15;2(3):100182. doi: 10.1016/j.xops.2022.100182

Table 3.

Subgroup Analysis of Experimental Conditions versus Real-World Conditions in Agreement and Correlation between the NYU Langone Eye Test Application and the Rosenbaum Card

Device by Conditions Agreement with Rosenbaum Card (Mean ± 95% Limits of Agreement, logMAR) Correlation with Rosenbaum Card (Intraclass Correlation Coefficient)
Experimental
conditions (50 participants)
 iPhone 0.008 ± 0.16 0.90
 Android 0.001 ± 0.13 0.92
Real-world
conditions (75 participants)
 iPhone 0.024 ± 0.35 0.66
 Android 0.016 ± 0.36 0.63

logMAR = logarithm of the minimum angle of resolution; NYU = New York University.