Accuracy (A) and precision (P) for the phantom measurements, stratified on environmental and system conditions, with the number of measurement points (n)

n Density Volume
A [pp] P [pp] A [%] P [%]
Normal conditions 50 0.6 1.1 1.8 1.2
Temperature variation 50 0.5 1.3 1.8 1.5
High load 25 −1.3 2.3 3.2 1.2
Old x‐ray tube 25 0.5 1.3 1.7 1.7
High density 12 −0.7 3.3 0.4 1.8
Total 162 0.2 1.8 1.9 1.6