Skip to main content
. 1998 Jul;5(4):531–536. doi: 10.1128/cdli.5.4.531-536.1998

TABLE 1.

Effect of adding NMS to sample diluent to remove false reactivity of plasma in the human IFN-γ EIA

Type of normal human plasmaa Level of added human IFN-γ (IU/ml) EIA optical density (450 nm)
No NMS 20% NMS
Nonreactive 0 0.069 0.062
50 1.475 1.419
Falsely reactive 0 2.548 0.063
50 2.735 1.422
a

Plasma from unstimulated blood donation.