Skip to main content
. 2021 Jun 26;1(2):100034. doi: 10.1016/j.xops.2021.100034

Table 3.

Agreement between the Notal OCT Analyzer and 2 Human Expert Graders for the Segmentation of Retinal Fluid at the Pixel Level on B-Scans

Grader Median Dice Coefficient
Total Retinal Fluid (%)
Intraretinal Fluid (%)
Subretinal Fluid (%)
Grader 1 Grader 2 Notal OCT Analyzer Grader 1 Grader 2 Notal OCT Analyzer Grader 1 Grader 2 Notal OCT Analyzer
Grader 1 71.2 72.6 72.4 75.9 70.6 70.3
Grader 2 71.2 73.5 72.4 77.8 70.6 70.9
Notal OCT Analyzer 72.6 73.5 75.9 77.8 70.3 70.9

— = not applicable.