Table 6.

Diagnostic accuracy, DOR, sensitivity and specificity for every examination technique.

Examination Technique Diagnostic Accuracy (%) Diagnostic Odds Ratio Sens. (%) Spec. (%)
B-mode blurry edge demarcation 83.6 2.5 75.0 63.7
microcaclifications 83.6 9.4 60.0 86.3
inhomogeneous sonomorphological structure 83.6 4.7 80.0 53.9
marginal hypoechogenicity 83.6 11.8 95.0 46.1
central hypoechogenicity 83.6 5.5 90.0 41.2
Shear-wave
elasto-graphy
marginal cut-off: 4.0 m/s 90.0 3.5 75.0 81.4
central cut-off: 3.9 m/s 88.4 2.3 75.0 70.1
marginal cut-off: 50.7 kPa 90.8 1.1 75.0 75.3
central cut-off: 46.9 kPa 89.2 1.0 75.0 72.2
CEUS fast marginal wash-in 88.7 21.3 50.0 96.9
fast central wash-in 82.6 0.9 20.0 96.8
inhomogeneous wash-in 82.6 7.2 34.9 93.1
partial/complete wash-out 82.6 3.4 66.7 95.2
CEUS: TIC marginal TTP cut-off: 13.1 s 78.5 1.1 57.9 81.1
central TTP cut-off: 13.5 s 78.5 1.1 78.9 73.0