TABLE 1.
Differences due to reagent sources in cytokine and soluble activation marker levels in plasma from HIV-positive subjects
Analyte (units of concn) | Commercial kit (type) from: | n | Mean level ± SD | P value (paired t test) | Pearson’s correlation coefficient |
---|---|---|---|---|---|
β2M (mg/liter) | Pharmacia (EIA) | 14 | 3.00 ± 1.25 | 0.0004a | 0.66a |
Abbott (MEIA) | 1.91 ± 0.70 | 0.0099b | 0.58b | ||
ORGenTec (EIA) | 2.25 ± 1.00 | 0.0085c | 0.91c | ||
Neopterin (nmol/liter) | ICN (EIA) | 17 | 28.1 ± 13.2 | 0.00001 | 0.95 |
BRAHMS (RIA) | 22.2 ± 10.9 | ||||
sIL-2R (pg/ml) | Endogen (EIA) | 20 | 8,708 ± 6,118 | 0.00002 | 0.58 |
Immunotech (EIA) | 2,205 ± 1,295 | ||||
sTNF-RII (ng/ml) | HyCult (EIA) | 42 | 4.59 ± 1.59 | 0.00001 | 0.96 |
Genzyme (EIA) | 3.82 ± 1.97 | ||||
TNF-α (pg/ml) | Innogenetics (EIA) | 35 | 335.8 ± 435 | 0.1032 | 0.91 |
Medgenix (EIA) | 398.0 ± 638 | ||||
IFN-γ (units/liter) | Immunotech (EIA) | 12 | 532 ± 598 | 0.0166 | 0.98 |
T-Cell Diagnostics (EIA) | 452 ± 532 | ||||
IL-10 (pg/ml) | Immunotech (EIA) | 23 | 49.8 ± 30.1 | 0.00004 | 0.03 |
Endogen (EIA) | 12.6 ± 9.3 |
Pharmacia β2M versus Abbott β2M.
Pharmacia β2M versus ORGentec β2M.
ORGentec β2M versus Abbott β2M.