Table 1.

Helmet settings in patients with acute hypoxemic respiratory, with and without pressure support

Ventilatory setting Helmet NIV Helmet CPAP
Ventilatory circuit Ventilator with bitube circuit and antimicrobial filter on expiratory port High flow generator with PEEP valve and antimicrobial filter on expiratory port
PEEP 10–15 cmH2O 10–15 cmH2O
Pressure support 10–14 cmH2O
Fresh gas flow 50–60 L/min
FiO2 Titrated to obtain SpO2 ≥ 92% and ≤ 98% Titrated to obtain SpO2 ≥ 92% and ≤ 98%
Pressurization rate 0.00 s (or fastest possible pressurization rate)
Inspiratory flow trigger 2 L/min or 2 cmH2O
Cycling 10–50% of maximum inspiratory flow
Maximum inspiratory time 1.2 s
Gas conditioning No humidification needed if minute ventilation < 35 L/min Active heating and humidification (37 °C or 34 °C according to patient’s comfort)