Duke Centered | Ella Centered | Charlie Centered | Charlie –50 mm FH | Charlie +50 mm FH | |
---|---|---|---|---|---|
Global SAR | |||||
Average SAR for 1 W input power (W/kg) | 0.066 | 0.069 | 0.091 | 0.075 | 0.109 |
Average SAR per B12 (W/kg/μT2) | 0.462 | 0.465 | 0.320 | 0.477 | 0.484 |
Average B1+ in central section for 1 W input power (μT)a | 0.379 | 0.385 | 0.542 | 0.397 | 0.475 |
Peak SAR | |||||
Peak local SAR (10 g averaged) for 1 W input power (W/kg) | 0.435 | 0.398 | 0.487 | 0.345 | 0.643 |
Peak local SAR (10 g averaged) per B12 (W/kg/μT2) | 3.04 | 2.63 | 1.72 | 2.19 | 2.85 |
The power optimization procedure of the MR imaging scanner software calibrates the needed input power to achieve a certain B1 in the subject. This calibration is based on the average B1+ in a central section of the subject (brain in this case).