Skip to main content
. 2022 Oct;43(10):E41–E42. doi: 10.3174/ajnr.A7660
Duke Centered Ella Centered Charlie Centered Charlie –50 mm FH Charlie +50 mm FH
Global SAR
 Average SAR for 1 W input power (W/kg) 0.066 0.069 0.091 0.075 0.109
 Average SAR per B12 (W/kg/μT2) 0.462 0.465 0.320 0.477 0.484
 Average B1+ in central section for 1 W input power (μT)a 0.379 0.385 0.542 0.397 0.475
Peak SAR
 Peak local SAR (10 g averaged) for 1 W input power (W/kg) 0.435 0.398 0.487 0.345 0.643
 Peak local SAR (10 g averaged) per B12 (W/kg/μT2) 3.04 2.63 1.72 2.19 2.85
a

The power optimization procedure of the MR imaging scanner software calibrates the needed input power to achieve a certain B1 in the subject. This calibration is based on the average B1+ in a central section of the subject (brain in this case).