Skip to main content
. 2022 Oct 17;1212:123510. doi: 10.1016/j.jchromb.2022.123510

Table 2.

Inter-day precision and accuracy (n = 6).

Sample
Nirmatrelvir
Ritonavir

Spike
Amount detected
Accuracy*
CV**
Spike
Amount detected
Accuracy*
CV**
(ng/mL) (ng/mL) (%) (%) (ng/mL) (ng/mL) (%) (%)
LLOQ 10 8.95 −10.5 7.31 2 1.68 −16.2 6.40
QC-low 20 21.8 9.19 9.46 4 3.63 −9.28 4.31
QC-med 500 536 7.30 5.56 100 102.1 2.14 3.48
QC-high 6000 6222 3.70 4.48 1200 1270 5.87 2.84
*

Accuracy: relative difference from expected value.

**

CV: coefficient of variation.