Skip to main content
. Author manuscript; available in PMC: 2023 Jul 1.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2022 Jun 30;69(7):2257–2269. doi: 10.1109/TUFFC.2022.3172690

TABLE III.

Results from assessing the combination of filters. The best filters are shown for multiple positive condition rates (15, 20, and 25% correspond to filter sets A-C). Neither the maximum brightness percentile nor the maximum standard deviation percentile contributed to any of the best-case options.

Filter Set Positive Predictive Value Positive Condition Rate Minimum Brightness Percentile Minimum Speckle SNR Minimum Frame-to-frame Correlation Maximum Standard Deviation Percentile Maximum Brightness Percentile
A 98.59% 15.06% 64 1.45 0.5 100 (off) 100 (off)
B 98.10% 20.76% 46 1.40 0.6 100 (off) 100 (off)
C 97.37% 25.39% 44 1.45 0.4 100 (off) 100 (off)