Table 2.

LC gradient condition

Column ACQUITY UPLC BEH C18 column (1.7 μm, 2.1 × 50 mm), ACQUITY UPLC BEH C18 vanguard pre-column (1.7 μm, 2.1 × 5 mm)
Column temperature 45°C
Injecting volume 5.0 μL
Flow rate 0.15 mL/min

Time (min) Gradient B %

0 20.0
1.0 20.0
3.0 60.0
20.0 98.0
20.1 20.0
22.0 20.0
25.0 20.0

Diverter time program

Time (min) Position

0.0 B
3.0 A
15.0 B

See also Table S1.