Skip to main content
. 2000 May;7(3):344–351. doi: 10.1128/cdli.7.3.344-351.2000

FIG. 3.

FIG. 3

Source of absolute CD4 T-cell count bias in individual laboratories. The absolute CD4 T-cell count generated in each laboratory by the predicate or tetraONE method was subtracted from the five-laboratory median value by that method to determine whether method biases occurred in individual laboratories. A through E are laboratory designations. The shaded region indicates the interquartile range (25th and 75th percentiles) of these differences.