Skip to main content
. 2000 May;7(3):344–351. doi: 10.1128/cdli.7.3.344-351.2000

TABLE 5.

Comparison of within-laboratory precision of CD4 and CD8 T-cell percentages for predicate method versus Flow-Count and tetraONE methodsa

T cells CD4 stratum (cells/μl) Median %CV (P vs predicate method)
Predicate method Flow-Count method tetraONE method
CD4 All 4.1 4.5 (0.530) 3.7 (0.001)
<200 6.5 6.0 (0.863) 4.3 (0.001)
>200 3.4 3.3 (0.240) 3.0 (0.225)
CD8 All 1.9 1.8 (0.515) 1.8 (0.268)
<200 1.8 2.0 (0.697) 1.5 (0.016)
>200 1.9 1.8 (0.565) 2.0 (0.565)
a

See Table 4, footnote a