Skip to main content
. Author manuscript; available in PMC: 2023 Nov 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2022 Aug 27;1210:123437. doi: 10.1016/j.jchromb.2022.123437

Table 3.

Matrix effect and extraction recovery of GCDCA-S and CDCA in different matrices.

Conc. (ng/mL) QC Matrices Matrix Effect Recovery
Mean Matrix Effect (%) CV (%) Mean Recovery (%) CV (%)
GCDCA-S
1.5 Pooled mouse plasma 111 10.9 88.8 9.58
90 115a 11.2 120 13.6
5 pooled human plasma NAb NAb 80.7 1.16
900 111 6.70 114 6.72
CDCA-24G
1.5 Pooled mouse plasma 88.0 8.88 89.8 7.26
90 109 8.70 104 3.78
1.5 pooled human plasma NAb NAb 90.6 12.2
90 87.1 11.1 94.0 4.91
a

The calculated mean matrix effect 114.814 was rounded to 115 as 3 significant figures kept.

b

The matrix effect result of prepared GCDCA-S and CDCA-24G at 5 ng/mL and 1.5 ng/mL in pooled human plasma was not available.