Table 4.

Stability of GCDCA-S and CDCA-24G in different matrices under various conditions.

Conc. (ng/mL) QC Matrices Sample Conditions
Bench-Top Stabilitya Long-Term Stabilityb Re-injection Stabilityc Freeze-Thaw Stabilityd
Mean Deviation (%) of t = 0 CV (%) Mean Deviation (%) of t = 0 CV (%) Mean Deviation (%) CV (%) Mean Deviation (%) of nominal Conc. CV (%)
GCDCA-S
1.5 Pooled mouse plasma 5.25 4.75 1.94 3.45 −6.79 4.53 7.47 5.26
90 −7.22 5.55 7.48 4.00 6.00 5.38 −1.51 4.50
5 water −5.61 8.06 2.45 7.14 −8.33 6.70 2.93 3.08
900 pooled human plasma −1.54 3.68 7.75 7.23 2.16 4.59 −7.09 4.66
CDCA-24G
1.5 Pooled mouse plasma −7.20 5.82 −8.00 5.95 1.69 3.35 −5.53 4.77
90 −7.57 8.37 −5.14 6.40 −0.0600 7.20 −14.3 9.14
1.5 water −2.05 4.17 −3.12 4.94 −4.34 4.06 −5.78 5.26
90 pooled human plasma −4.68 4.00 1.94 2.63 −6.50 5.85 −4.45 3.96
a

Stored at ambient temperature (25 °C) for 6 hours,

b

Stored for 28 days at −80°C,

c

Stored the analyzed samples at auto-sampler (4°C) and reanalyzed samples after 24 hours,

d

After three freeze-thaw cycles.