Table 1.
Percentage depth dose measured/estimated from parallel plate chamber (PPC), EBT3 films, and PRIMO MC (Monte Carlo code name) for a 6 MV photon beam for 0º gantry angle.
| Depth (mm) | PPC measured value | EBT3 Films measured value | PRIMO MC simulated results | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 5×5 | 10×10 | 15×15 | 20×20 | 5×5 | 10×10 | 15×15 | 20×20 | 5×5 | 10×10 | 15×15 | 20×20 | |
| 0 | 14.78 | 19.87 | 25.83 | 31.54 | 14.20 | 19.14 | 25.14 | 30.90 | 13.29 | 17.98 | 23.09 | 28.47 |
| 1 | 45.84 | 46.92 | 51.96 | 55.70 | 45.55 | 46.69 | 51.50 | 54.76 | 43.55 | 44.74 | 49.26 | 52.92 |
| 2 | 61.96 | 63.24 | 68.13 | 71.50 | 61.21 | 63.56 | 67.89 | 71.65 | 59.94 | 61.48 | 65.06 | 68.18 |
| 3 | 73.14 | 75.46 | 78.20 | 81.75 | 72.29 | 77.39 | 77.92 | 81.59 | 70.99 | 72.50 | 75.85 | 78.23 |
| 4 | 82.03 | 82.83 | 86.37 | 87.63 | 81.21 | 84.87 | 85.41 | 87.56 | 79.21 | 80.98 | 83.19 | 85.27 |
| 5 | 86.83 | 88.10 | 90.39 | 91.47 | 85.82 | 87.70 | 90.32 | 91.72 | 84.98 | 86.69 | 88.62 | 90.17 |
| 7 | 93.94 | 94.72 | 95.74 | 96.26 | 93.01 | 94.31 | 95.80 | 96.81 | 92.49 | 93.31 | 95.04 | 95.96 |
| 10 | 99.17 | 99.14 | 99.70 | 98.80 | 98.13 | 98.98 | 99.62 | 99.10 | 97.93 | 98.62 | 99.16 | 99.63 |
PPC: Parallel plate chamber, EBT3: Brand name, PRIMO MC: Monte Carlo code name