Skip to main content
. 2000 May;7(3):451–456. doi: 10.1128/cdli.7.3.451-456.2000

TABLE 5.

Discordant results in tests for heterophile antibodies for patients with recent primary EBV infection (group A)

Serum no. Result with:
Paul-Bunnell-Davidsohn test Monosticon Avitex Alexon Clearview Cards±OS
130 +
131 + + + +
132 + + + + +
133 +/−
134 + + + + +
135 + + + + +
139 +/− +
140 +/−
141 + + +/− +
176 + + + + +
178 + + + + +
190 + + + + +
191 + + +
192 + + +
213 + + + + +
214 + + + + +
217 + + + + +
226 + + + +
245 (+)a
246 +/− + + + +
250 + + + +/−
251 + + + + +
257 + + + + +
258 + + + + +
271 + + + + +
291 + + + +/− +
292 + + + + +
296 + + + + +
297 + + + + +
300 + + + +
304 + +/− +/− +
323 + + + (+)a +
336 +
a

A positive reaction appeared outside the time specified by the manufacturer.