Table 4.

Application of the proposed sensor Au@TiO2/MWCNTs/GCE for real sample analysis (human serum and saliva samples).

Sample Added (µM) Found in Serum (µM)
(n = 3)
Recovery
(%)
Found in Saliva (µM)
(n = 3)
Recovery
(%)
1 20 18.02 ± 0.94 90.09 18.33 ± 0.71 91.63
40 38.10 ± 0.89 95.24 38.28 ± 0.60 95.71
60 51.17 ± 0.70 85.29 70.74 ± 1.43 117.90
2 20 20.70 ± 1.20 103.48 21.30 ± 0.87 106.52
40 37.41 ± 0.64 93.53 43.98 ± 0.69 109.95
60 57.39 ± 0.56 95.65 63.85 ± 0.45 106.41
3 20 22.11 ± 0.88 110.54 22.76 ± 1.02 113.80
40 38.10 ± 1.09 95.24 39.63 ± 0.70 99.08
60 51.17 ± 0.79 85.29 68.30 ± 0.83 113.84