TABLE 2.
Aptima bacterial vaginosis assay performance on Panther system |
||||
---|---|---|---|---|
Reference method: Nugent score (NS) | True positive samples | True negative samples | False positive samples | False negative samples |
NS ≤ 3 | 0 | 136 | 8 | 0 |
NS ≥ 7 | 41 | 0 | 0 | 4 |
Aptima bacterial vaginosis assay performance on Panther system |
||||
---|---|---|---|---|
Reference method: Nugent score (NS) | True positive samples | True negative samples | False positive samples | False negative samples |
NS ≤ 3 | 0 | 136 | 8 | 0 |
NS ≥ 7 | 41 | 0 | 0 | 4 |