. 2021 Nov 8;14(11):1139–1144. doi: 10.1136/neurintsurg-2021-018220

Table 2.

Occupational radiation doses in the hybrid operating room mapped along positions A-H (see figure 1B) for the hybrid cone beam CT and the O-arm cone beam CT systems

Distance (cm) Device A B C D E F G H
Hp(10) (µSv)
100 hCBCT 39.1±0.4 45.0±0.8 36.9±1.8 46.9±2.1 52.4±3.5
oCBCT 62.9±1.2 29.5±1.0 70.4±0.3 46.6±1.7 62.6±1.2 63.3±1.5
150 hCBCT 20.2±0.5 13.4±0.1 23.7±0.2 15.7±0.2 25.9±1.7 26.5±2.3
oCBCT 29.2±0.4 12.6±0.1 33.3±0.6 18.8±0.5 30.0±0.5 29.8±0.7
200 hCBCT 12.3±0.2 8.0±0.1 14.6±0.1 8.2±0.3 15.9±0.6 0.8±0.0 16.3±0.8 6.8±0.3
oCBCT 16.4±0.1 9.3±0.1 17.1±0.4 9.8±0.2 17.4±0.3 2.7±0.0 17.6±0.5 8.9±0.1
300 hCBCT 6.1±0.2 3.9±0.0 6.8±0.1 3.6±0.1 7.4±0.4 0.8±0.0 7.4±0.6 2.9±0.2
oCBCT 7.8±0.2 4.3±0.1 8.2±0.1 3.9±0.2 7.6±0.1 1.0±0.0 7.8±0.1 3.4±0.1

Results are mean±SD.

Some data are missing because of the operating room table or X-ray system obstructing the measurement position. The radiation doses were measured per rotational acquisition in terms of the personal dose equivalent, Hp(10) (µSv), using standard clinical imaging protocols (described in table 1).

hCBCT, hybrid cone beam CT; oCBCT, O-arm cone beam CT.