Skip to main content
. 2022 Sep 22;13(10):1573. doi: 10.3390/mi13101573

Table 2.

An overview of the checkpoints used to assess the outcome of different stages in the BBB chips timeline. These checkpoints include PDMS bonding, media perfusion, cell attachment and the formation of a confluent monolayer. Common and suspected failure modes are also described.

Description Common and Suspected Failure Modes
Sample size of chips between checkpoints 1–2 17
Checkpoint 1 Successful PDMS-PDMS bonding based on complete transparency of PDMS pieces and manual leak test 100% N/A
Checkpoint 2 Viable chips at the end of the experiment 88% Tubing coming undone from needle on syringe that is dispensing media.
Sample size of chips between checkpoints 3–4 8
Checkpoint 3 Chips with cell attachment 88% Cell detachment due to uneven ECM coating, air bubbles in the channel, or insufficient media supply.
Checkpoint 4 Chips with a confluent monolayer 57% Washing and fixing steps and disassembly of chips caused cells to lift before mounting membranes onto glass slides. Cell detachment due to uneven ECM coating.